ViCTor
"Value for money" eddy current flaw detector for component testing i on-line applications.
The ViCTor will more than match any competitor due to its outstanding signal-to-noise performance. Intended for automated applications such as component testing, the ViCTor perfectly suits its role. However, with the ability to team with a National Instruments LabView software development-kit, the ViCTor has the flexibility to be application-specific also.
It is the ViCTor’s ability to function with the LabView Development Kit that sets it apart. This functionality provides a simple engineer-friendly software solution to permit more advanced features to be added to customise system operation.
Alternatively, if an application only requires a limited range of features, the ViCTor has a user-friendly interface that can be edited further to suit even the most inexperienced operator.
CUSTOMER BENEFITS
- Ultra low noise
- Simple and easy to use interface
- Compatible with virtually any PC
- Industry leading test frequency range
- Powered by easy to find C cells
- Great value for money
APPLICATIONS
- Research
- Weld Inspection
- Training Schools
- Corrosion Assessment
- Surface Crack Detection and Assessment
- Sub-surface defect detection on non-ferrous material
Able to be used with virtually any PC, the ViCTor is both cost effective and has the flexibility to be optimised for specific applications.
The ViCTor is available in both a single-channel or dual—channel configuration. The ViCTor 2-Channel allows the operator to test an area at both low and high frequency simultaneously.
Specification
Probe | Modes | Selectable Bridge (including Absolute InBal probes) or Reflection Probes |
---|---|---|
Frequency | 2Hz-37MHz | |
Drive | 1-20 v pk in 0.1 v increments | |
Signal | Phase | 0°-359° in 1° increments |
Gain | 0-100dB in 1dB increments | |
Digitising | 8 kHz maximum | |
Filter | Low, High and Band Pass, eight Pole Bessel. Low pass filter: 10-500Hz in 1Hz increments, 500-2000Hz in 50Hz increments. High Pass: 0-500Hz in 1Hz increments. Band Pass addition of Low and High Pass. | |
Display | Mode | Phase (impedance) Plane |
Grid | 10 by 10 graticule | |
Spot Position | Separate X and Y Controls | |
Alarm | Type | Rectangular alarm |
Setting | Drag and Drop | |
Output | Open collector, LED and 80dBA Audio Alarm with volume control. | |
Storage | Screen | Save currently displayed screen settings for reports |
Setting | Save and recall of settings | |
Recording | Save recorded data to disk off line analysis | |
Computer | Type | PC running Windows XP™ or Windows 7™ |
Interface | USB 2.0 | |
Size and Weight | Dimensions | 12.7cm (5.0") Long x 8.0cm (3.2") Wide x 4.7cm (1.85") Thick |
Weight | 0.6kg (1.3lbs) | |
Power | DC | 5v at 1.2 A DC |
AC | With adapter 100-240 v AC 50-60Hz 0.3A |